Magnetometry with ensembles of nitrogen vacancy centers in bulk diamond
Author(s)
Teale, Carson (Carson Arthur)
DownloadFull printable version (7.052Mb)
Other Contributors
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science.
Advisor
Dirk Englund and Danielle Braje.
Terms of use
Metadata
Show full item recordAbstract
This thesis summarizes experiments conducted to develop a high sensitivity vector magnetometer using nitrogen vacancy (NV) centers in a bulk diamond sample. This project began by analyzing the sensitivity of a single NV orientation using a continuous wave electron spin resonance approach. A protocol for determining the diamond's orientation was developed to map vector magnetic field readings in the diamond reference frame to the lab frame. Preliminary vector field measurements and differential vector measurements were performed. Although these showed promising results, significant instrument and ambient magnetic noise limited the achievable sensitivity. A new frequency locking measurement technique was developed to allow for simultaneous measurements between two separate sensors for future differential experiments. This technique provides a host of other benefits including much improved dynamic range and steady-state immunity to fluctuations in linewidth and contrast.
Description
Thesis: S.M., Massachusetts Institute of Technology, Department of Electrical Engineering and Computer Science, 2015. Cataloged from PDF version of thesis. Includes bibliographical references (pages 55-57).
Date issued
2015Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer SciencePublisher
Massachusetts Institute of Technology
Keywords
Electrical Engineering and Computer Science.