dc.description.abstractFiber optical arrangements are useful for measurement of aquatic chemical species, especially in waters of high chemical concentrations. An optical fiber spectrofluorometer device known as the "Hammerhead" was developed in the Hemond lab in 2015 and improved in 2016 to measure concentrations of dissolved organic matter in water samples. The Hammerhead operates by detecting fluorescence, absorbance, and scattering of light within the water. However, inner shielding, particularly in high DOC concentrations, can lead to meaningless results. This project expands upon the earlier Hammerhead work by seeking the highest possible fluorescence signal from the Hammerhead. Experiments were run using a series of different geometric schemes for the Hammerhead optical fibers. Of the configurations tested, the narrowest spacing between excitation fiber and detector (0.076 inches from each fiber's tip to the center of the chamber) produced optimal results. This new configuration shortens the light path length and largely avoids inner shielding effects. The new design was then compared with a traditional flow cell water measurement device known as the LEDIF, using both fluorescein and humic acid substances. Comparing the signals for both instruments indicates that the Hammerhead fiber optic scheme is superior in detecting fluorescence measurements at high DOC concentrations; it produces a nearly linear response for fluorescence, while inner shielding hinders LEDIF results. In addition, comparing the limit of detection for both instruments indicates that the Hammerhead performs better than the LEDIF even independent of inner shielding effects, producing a signal nearly 10 times greater than the LEDIF in low DOC concentrations.en_US
dc.description.statementofresponsibilityby Abigail E. Cartwright.en_US
dc.format.extent92 pagesen_US
dc.language.isoengen_US
dc.publisherMassachusetts Institute of Technologyen_US
dc.rightsMIT theses are protected by copyright. They may be viewed, downloaded, or printed from this source but further reproduction or distribution in any format is prohibited without written permission.en_US
dc.rights.urihttp://dspace.mit.edu/handle/1721.1/7582en_US
dc.subjectCivil and Environmental Engineering.en_US
dc.titleOptimization of a fiber optic freshwater DOM sensing deviceen_US
dc.typeThesisen_US
dc.description.degreeM. Eng.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Civil and Environmental Engineeringen_US
dc.identifier.oclc1129584561en_US
dc.description.collectionM.Eng. Massachusetts Institute of Technology, Department of Civil and Environmental Engineeringen_US
dspace.imported2019-12-13T18:52:30Zen_US
mit.thesis.degreeMasteren_US
mit.thesis.departmentCivEngen_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record