EMG methods for prosthesis ankle-subtalar free-space control

Author(s)
Qiao, Junqing
Thumbnail
DownloadThesis PDF (99.52Mb)
Advisor
Herr, Hugh M.
Terms of use
In Copyright - Educational Use Permitted Copyright MIT http://rightsstatements.org/page/InC-EDU/1.0/
Metadata
Show full item record
Abstract
EMG-based prosthetic joint controllers have been an active research field for more than fifty years. However, several challenges remain to be addressed[9]. Electrodes positioning, controllers calibration, and controllers’ linear approximation error are the most challenging problems among them. This thesis introduces three methods to solve those problems respectively. They are 1. A non-negative blind source separation algorithm named non-negative orthogonal decomposition(NOD). This algorithm aims to replace non-negative matrix factorization(NMF) for muscle motion base extraction. NOD recovers the source signal by finding the borders of the input signal and translating the borders onto the coordinate axis. The translated signals are the recovered signals. 2. An unsupervised algorithm for generating joint trajectories from EMG signals in reciprocating movements. The EMG signal and trajectory can be used for EMG-based prosthesis joint controller calibration. And 3. an innovative EMG-to-joint-position controller. It uses neural networks to compensate for the nonlinearity of the well-known bilinear model[2]. The NOD algorithm successfully extracted motion bases from the EMG signals. Compared with NMF, the motion bases are more independent and stable. The minimum-jerk-based trajectory generator generated smooth and biomimetic trajectories on intact subjects. The trajectories are close to the ground truth collected from the goniometer. The third model also has considerable improvement in joint angle accuracy over the linear muscle model.
Date issued
2023-02
URI
https://hdl.handle.net/1721.1/152019
Department
Program in Media Arts and Sciences (Massachusetts Institute of Technology)
Publisher
Massachusetts Institute of Technology

Collections
  • Graduate Theses

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.