Show simple item record

dc.contributor.advisorHarry H. Asada and Boo-Ho Yang.en_US
dc.contributor.authorRhee, Sokwooen_US
dc.contributor.otherMassachusetts Institute of Technology. Dept. of Mechanical Engineering.en_US
dc.date.accessioned2006-05-15T20:23:41Z
dc.date.available2006-05-15T20:23:41Z
dc.date.copyright2000en_US
dc.date.issued2000en_US
dc.identifier.urihttp://hdl.handle.net/1721.1/32706
dc.descriptionThesis (Ph.D.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2000.en_US
dc.descriptionIncludes bibliographical references (leaves 98-101).en_US
dc.description.abstractA miniaturized, telemetric, photoplethysmograph sensor for long-term, continuous monitoring is presented in this thesis. The sensor, called a "ring sensor," is attached to a finger base for monitoring beat-to-beat pulsation, and the data is sent to a host computer via a RF transmitter. Two major design issues are addressed: one is to minimize motion artifact and the other is to minimize the consumption of battery power. An efficient double ring design is developed to lower the influence of external force, acceleration, and ambient light, and to hold the sensor gently and securely on the skin, so that the circulation at the finger may not be obstructed. To better understand the mechanism of motion artifact by external forces, a comprehensive mathematical model describing the finger photoplethysmography was developed and verified by finite element method, numerical simulation and experiments. Total power consumption is analyzed in relation to the characteristics of the individual components, sampling rate, and CPU clock speed. Optimal operating conditions are obtained for minimizing the power budget. A prototype ring sensor is designed and built based on the power budget analysis and the artifact-resistive attachment method.en_US
dc.description.abstract(cont.) It is verified through experiments that the ring sensor is resistant to interfering forces and acceleration acting on the ring body. It is also shown that the device meets diverse and conflicting requirements, including compactness, motion artifact reduction, minimum loading effects, and low battery power consumption. Benchmarking tests with FDA-approved photoplethysmograph and EKG reveal that the ring sensor is comparable to those devices in detecting beat-to-beat pulsation despite disturbances. The long-term monitoring experiment shows that this device can effectively provide a considerable amount of artifact-free vital sign information in everyday life. Finally, guidelines for designing the ring sensor are proposed based on the analyses and the experiment results.en_US
dc.description.statementofresponsibilityby Sokwoo Rhee.en_US
dc.format.extent101 leavesen_US
dc.format.extent6681174 bytes
dc.format.extent6686521 bytes
dc.format.mimetypeapplication/pdf
dc.format.mimetypeapplication/pdf
dc.language.isoengen_US
dc.publisherMassachusetts Institute of Technologyen_US
dc.rightsM.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.en_US
dc.rights.urihttp://dspace.mit.edu/handle/1721.1/7582
dc.subjectMechanical Engineering.en_US
dc.titleDesign and analysis of artifact-resistive finger photoplethysmographic sensors for vital sign monitoringen_US
dc.typeThesisen_US
dc.description.degreePh.D.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Mechanical Engineering
dc.identifier.oclc48747324en_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record