Show simple item record

dc.contributor.advisorKamal Youcef-Toumi.en_US
dc.contributor.authorEl Rifai, Osamah Men_US
dc.contributor.otherMassachusetts Institute of Technology. Dept. of Mechanical Engineering.en_US
dc.date.accessioned2007-08-03T18:21:04Z
dc.date.available2007-08-03T18:21:04Z
dc.date.copyright2002en_US
dc.date.issued2002en_US
dc.identifier.urihttp://hdl.handle.net/1721.1/38256
dc.descriptionThesis (Ph. D.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2002.en_US
dc.descriptionIncludes bibliographical references (leaves 156-165).en_US
dc.description.abstractThe phenomenal resolution and versatility of the atomic force microscope (AFM), has made it a widely-used instrument in nanotechnology. In this thesis, a detailed model of AFM dynamics has been developed. It includes a new model for the piezoelectric scanner coupled longitudinal and lateral dynamics, creep, and hysteresis. Models for probe-sample interactions and cantilever dynamics were also included. The models were used to improve the dynamic response and hence image quality of contact-mode AFM. An extensive parametric study has been performed to experimentally analyze in-contact dynamics. Nonlinear variations in the frequency response were observed, in addition to changes in the pole-zero structure. The choice of scan parameters was found to have a major impact on image quality and feedback performance. Further, compensation for scanner creep was experimentally tested yielding a reduction in creep by a factor of 3 to 4 from the uncompensated system. Moreover, fundamental performance limitations in the AFM feedback system were identified. These limitations resulted in a severe bound on the maximum achievable feedback bandwidth, as well as a fundamental trade-off between step response overshoot and response time. A careful analysis has revealed that a PID controller has no real advantage over an integral controller.en_US
dc.description.abstract(cont.) Therefore, a procedure for automatically selecting key scan parameters and controller gain was developed and experimentally tested for I-control. This approach, in contrast to the commonly used trial and error method, can substantially improve image quality and fidelity. In addition, a robust adaptive output controller (RAOC), was designed to guarantee global boundedness and asymptotic regulation in the presence and absence of disturbances, respectively. Simulations have shown that a substantial reduction in contact force can be achieved with the RAOC, in comparison with a well-tuned I-controller, yet with no increase in the maximum scan speed. Furthermore, a new method was developed to allow calibrating the scanner's vertical displacement up to its full range, in addition to characterizing scanner hysteresis. This work has identified and addressed crucial problems and proposed practical solutions to factors limiting the dynamic performance of the AFM.en_US
dc.description.statementofresponsibilityby Osamah M. El Rifai.en_US
dc.format.extent167 leavesen_US
dc.language.isoengen_US
dc.publisherMassachusetts Institute of Technologyen_US
dc.rightsM.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.en_US
dc.rights.urihttp://dspace.mit.edu/handle/1721.1/7582
dc.subjectMechanical Engineering.en_US
dc.titleModeling and control of undesirable dynamics in atomic force microscopesen_US
dc.typeThesisen_US
dc.description.degreePh.D.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Mechanical Engineering
dc.identifier.oclc150951623en_US