Alcator C-Mod soft X-ray pulse height analysis system
Author(s)
Gamboa, Eliseo (Eliseo J.)
DownloadFull printable version (2.679Mb)
Other Contributors
Massachusetts Institute of Technology. Dept. of Physics.
Advisor
John E. Rice.
Terms of use
Metadata
Show full item recordAbstract
A pulse height analysis (PHA) system has been installed on the Alcator C-Mod magnetic confinement fusion experiment. The PHA utilizes a Si(Li) detector to measure soft X-rays in the 1-30 keV range with an energy resolution on the order of 100eV. A FASTComtec MCA-3 multichannel analyzer allows integration times of ims while the PHA supports counting rates of up to 80kHz. The thermal electron temperature is measured from the energy dependence of the intensity of the electron bremsstrahlung spectrum. Measurements of line radiation allows the PHA to identify impurity ions within each plasma shot. The result is a diagnostic that measures thermal electron temperatures while aiding in the indentification of heavy (Z 12) impurities.
Description
Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Physics, 2007. Includes bibliographical references (leaves 59-60).
Date issued
2007Department
Massachusetts Institute of Technology. Department of PhysicsPublisher
Massachusetts Institute of Technology
Keywords
Physics.