dc.contributor.advisor | Peter Fisher. | en_US |
dc.contributor.author | Chonde, Daniel Burje | en_US |
dc.contributor.other | Massachusetts Institute of Technology. Dept. of Physics. | en_US |
dc.date.accessioned | 2008-03-27T18:21:48Z | |
dc.date.available | 2008-03-27T18:21:48Z | |
dc.date.copyright | 2007 | en_US |
dc.date.issued | 2007 | en_US |
dc.identifier.uri | http://hdl.handle.net/1721.1/40905 | |
dc.description | Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Physics, 2007. | en_US |
dc.description | Includes bibliographical references (p. 53-54). | en_US |
dc.description.abstract | This paper examines a new technique for measuring gas composition through the use of a spirometer. A spirometer is high precision pressure transducer which measures the speed of sound in a gas through the emission and reception of ultrasonic pulses, commonly used in medicine to measure patient lung capacity. The spirometer was successfully calibrated to measure gas composition to an accuracy of 1.75% ± 1.23% by ratio by weight. By using a spirometer, the speed of sound was measured for a trapped volume of Argon CO₂ mixtures at ratios of 80 : 20, 85 : 15, 90 : 10, 95 : 5, and 100 : 0. The temperature dependence of the transit time was observed to behave according to the Laplace Formula, except for the presence of dramatic drops that were dependent on the Argon concentration. An argument is presented that these drops are technical limitations of the spirometer. | en_US |
dc.description.statementofresponsibility | by Daniel Burje Chonde. | en_US |
dc.format.extent | 54 p. | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Massachusetts Institute of Technology | en_US |
dc.rights | M.I.T. theses are protected by
copyright. They may be viewed from this source for any purpose, but
reproduction or distribution in any format is prohibited without written
permission. See provided URL for inquiries about permission. | en_US |
dc.rights.uri | http://dspace.mit.edu/handle/1721.1/7582 | en_US |
dc.subject | Physics. | en_US |
dc.title | Spirometer techniques for measuring molar composition in argon carbon dioxide mixtures | en_US |
dc.type | Thesis | en_US |
dc.description.degree | S.B. | en_US |
dc.contributor.department | Massachusetts Institute of Technology. Department of Physics | |
dc.identifier.oclc | 212374677 | en_US |