dc.description.abstract | This thesis details a measurement setup and experimental procedures for emittance measurements using a Fourier transform infrared spectrometer. We calibrate the FTIR measurement system using measurements of a blackbody source at two temperatures to obtain the true emittance and absolute spectral radiance of our samples. | en_US |
dc.description.statementofresponsibility | by Susannah Brown. | en_US |
dc.format.extent | 51 leaves | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Massachusetts Institute of Technology | en_US |
dc.rights | M.I.T. theses are protected by
copyright. They may be viewed from this source for any purpose, but
reproduction or distribution in any format is prohibited without written
permission. See provided URL for inquiries about permission. | en_US |
dc.rights.uri | http://dspace.mit.edu/handle/1721.1/7582 | en_US |
dc.subject | Physics. | en_US |
dc.title | Thermal emission measurement and calibration | en_US |
dc.type | Thesis | en_US |
dc.description.degree | S.B. | en_US |
dc.contributor.department | Massachusetts Institute of Technology. Department of Physics | |
dc.identifier.oclc | 495709760 | en_US |