Comparison of Spectral/Fourier Domain Optical Coherence Tomography Instruments for Assessment of Normal Macular Thickness
Author(s)
Sull, Alan C.; Vuong, Laurel N.; Price, Lori Lyn; Srinivasan, Vivek J.; Gorczynska, Iwona; Fujimoto, James G.; Schuman, Joel S.; Duker, Jay S.; ... Show more Show less
DownloadFujimoto-Comparison of spectral.pdf (2.208Mb)
OPEN_ACCESS_POLICY
Open Access Policy
Creative Commons Attribution-Noncommercial-Share Alike
Terms of use
Metadata
Show full item recordAbstract
Purpose: The purpose of this study was to report normal macular thickness measurements and assess reproducibility of retinal thickness measurements acquired by a time-domain optical coherence tomography (OCT) (Stratus, Carl Zeiss Meditec, Inc., Dublin, CA) and three commercially available spectral/Fourier domain OCT instruments (Cirrus HD-OCT, Carl Zeiss Meditec, Inc.; RTVue-100, Optovue, Inc., Fremont, CA; 3D OCT-1000, Topcon, Inc., Paramus, NJ).
Methods: Forty randomly selected eyes of 40 normal, healthy volunteers were imaged. Subjects were scanned twice during 1 visit and a subset of 25 was scanned again within 8 weeks. Retinal thickness measurements were automatically generated by OCT software and recorded after manual correction. Regression and Bland-Altman plots were used to compare agreement between instruments. Reproducibility was analyzed by using intraclass correlation coefficients, and incidence of artifacts was determined.
Results: Macular thickness measurements were found to have high reproducibility across all instruments with intraclass correlation coefficients values ranging 84.8% to 94.9% for Stratus OCT, 92.6% to 97.3% for Cirrus Cube, 76.4% to 93.7% for RTVue MM5, 61.1% to 96.8% for MM6, 93.1% to 97.9% for 3D OCT-1000 radial, and 31.5% to 97.5% for 3D macular scans. Incidence of artifacts was higher in spectral/Fourier domain instruments, ranging from 28.75% to 53.16%, compared with 17.46% in Stratus OCT. No significant age or sex trends were found in the measurements.
Conclusion: Commercial spectral/Fourier domain OCT instruments provide higher speed and axial resolution than the Stratus OCT, although they vary greatly in scanning protocols and are currently limited in their analysis functions. Further development of segmentation algorithms and quantitative features are needed to assist clinicians in objective use of these newer instruments to manage diseases.
Description
Author Manuscript 2011 February 1.
Date issued
2010-02Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer ScienceJournal
Retina
Publisher
Lippincott Williams & Wilkins
Citation
Alan C. Sull et al. "Comparison of Spectral/Fourier Domain Optical Coherence Tomography Instruments for Assessment of Normal Macular Thickness" Retina 30(2):235-245, February 2010.
Version: Author's final manuscript
ISSN
0275-004X