Development of a combined multi-sensor/signal processing architecture for improved in-situ quantification of the charge balance of natural waters
Author(s)
Mueller, Amy, 1980-
DownloadFull printable version (21.88Mb)
Other Contributors
Massachusetts Institute of Technology. Dept. of Civil and Environmental Engineering.
Advisor
Harold F. Hemond.
Terms of use
Metadata
Show full item recordAbstract
This thesis details the design, implementation, and testing of a new electrochemical instrument for the in situ measurement of both major and environmentally relevant minor ions in fresh waters, namely Na+, K+, Ca2+, Mg2+, NH+4 , Cl-, NO-3 , and SO2-4 . The instrument is built on a hybrid multi-probe / signal processing architecture and is implemented using commercial sensor hardware (primarily ion-selective electrodes (ISEs)) paired with a novel neural network processor designed to take advantage of a priori chemical knowledge about the system. Adaptation of this architecture to in-situ conditions and quantification of relatively minor ions required overcoming a number of challenges, including: (1) lack of a standardized method for unsupervised recording of ISE equilibrium potential, (2) non-availability of commercial electrodes for some ion species, and (3) detection of ion concentrations that fall below the ISE linear response region and/or are confounded by the presence of relatively large quantities of interfering ions. As such, a methodology is proposed and validated for standardization of ISE potential readings, resulting in consistent measurements completed in <6.5 min., improving replicability, and facilitating simultaneous measurement of up to 12 ion channels. The sensor suite is then designed such that each ISE provides information about more than one analyte, and finally, the artificial neural network (ANN) architecture is optimized for use on environmental chemical data by including software constraints implementing known chemical relationships, i.e., the concept of charge balance and the total ion-conductivity relationship. Two experiments are conducted using environmentally-relevant data sets (one semi-synthetic, one created in the lab) to characterize the eectiveness of the proposed ANN architecture. Final results demonstrate over an order of magnitude decrease in relative error (as measured against use of ISEs as stand-alone sensors) without concentration-dependent error bias, including estimates for analytes for which no specfic ISE exists (SO2-4 , Mg2+, HCO-3 ). Simultaneous un-biased quantification of all eight ions is achieved with ~20% error on most channels including NO-3 (concentrations </-100 [mu]M) and ~50% error for NH+4 (</-100 [mu]M), however it is also demonstrated that errors of ~10% are achievable for N-species ions even at low concentration if slightly higher uncertainties on other channels can be tolerated.
Description
Thesis (Ph. D.)--Massachusetts Institute of Technology, Dept. of Civil and Environmental Engineering, 2012. This electronic version was submitted by the student author. The certified thesis is available in the Institute Archives and Special Collections. Cataloged from student submitted PDF version of thesis. Includes bibliographical references (p. 189-197).
Date issued
2012Department
Massachusetts Institute of Technology. Department of Civil and Environmental EngineeringPublisher
Massachusetts Institute of Technology
Keywords
Civil and Environmental Engineering.