Show simple item record

dc.contributor.advisorKamal Youcef-Toumi.en_US
dc.contributor.authorAumond, Bernardo Dantas, 1972-en_US
dc.contributor.otherMassachusetts Institute of Technology. Department of Mechanical Engineering.en_US
dc.date.accessioned2014-08-19T17:35:55Z
dc.date.available2014-08-19T17:35:55Z
dc.date.copyright2001en_US
dc.date.issued2001en_US
dc.identifier.urihttp://hdl.handle.net/1721.1/88892
dc.descriptionThesis (Ph.D.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2001.en_US
dc.descriptionIncludes bibliographical references (leaves 186-190).en_US
dc.description.abstractMetrological data from sample surfaces can be obtained by using a variety of profilome try methods. Atomic Force Microscopy (AFM), which relies on contact inter-atomic forces to extract topographical images of a sample, is one such method that can be used on a wide range of surface types, with possible nanometer resolution (both vertical andlateral). However, AFM images are commonly distorted by convolution, which reduces metrological accuracy. This type of distortion is more significant when the sample surface containshigh aspect ratio features such as lines, steps or sharp edges or when probe and sample share similar characteristic dimensions. Therefore, as the size of engineered features arepushed into the micrometer and sub-micrometer range by the development of new high precision fabrication techniques, convolution distortions embedded in the images becomeincreasingly more significant. Aiming at mitigating these distortions and recovering metrology sound ness, we introduce a novel image deconvolution scheme based on the principle of stereo imaging. Multiple images of a sample, taken at different angles, allow for separation ofcon volution artifacts from true topographic data. As a result, accurate samplereconstruction and probe shape estimation can be achieved simultaneously. Additionally, shadow zones, which are areas of the sample that cannot be reached by the AFM probe, are greatly re duced. Most importantly, this technique does not require a priori probe characterizationor any sort of shape assumption. It also reduces the need for slender or sharper probes,which, on one hand, induce less convolution distortion but, on the other hand, are more prone to wear and damage, thus decreasing the overall inspection system reliability.en_US
dc.description.abstract(cont.) This research project includes a survey of current high precision metrology tools and an in-depthanalysis of the state of the art deconvolution techniques for probe based metrology instruments. Next, the stereo imaging algorithm is introduced, simulation results presented and anerror analysis is conducted. Finally, experimental validations of the technique are carried outfor an industrial inspection application where the characteristic dimensions of the samplesare in the nanometer range. The technique was found to be robust and insensitive to probe or shape geometries. Furthermore, the same framework was deemed to be applicable to other probe based imaging techniques such as mechanical stylus profilometers and scanning tunneling microscopy.en_US
dc.description.statementofresponsibilityby Bernardo Dantas Aumond.en_US
dc.format.extent206 leavesen_US
dc.language.isoengen_US
dc.publisherMassachusetts Institute of Technologyen_US
dc.rightsM.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.en_US
dc.rights.urihttp://dspace.mit.edu/handle/1721.1/7582en_US
dc.subjectMechanical Engineering.en_US
dc.titleHigh precision stereo profilometryen_US
dc.typeThesisen_US
dc.description.degreePh.D.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Mechanical Engineering
dc.identifier.oclc48925666en_US