<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-09-20T00:43:04Z</responseDate><request verb="GetRecord" identifier="oai:dspace.mit.edu:1721.1/18026" metadataPrefix="dim">https://dspace.mit.edu/server/oai/request</request><GetRecord><record><header><identifier>oai:dspace.mit.edu:1721.1/18026</identifier><datestamp>2022-01-13T07:54:29Z</datestamp><setSpec>com_1721.1_7582</setSpec><setSpec>com_1721.1_7581</setSpec><setSpec>col_1721.1_131023</setSpec></header><metadata><dim:dim xmlns:dim="http://www.dspace.org/xmlns/dspace/dim" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.dspace.org/xmlns/dspace/dim http://www.dspace.org/schema/dim.xsd">
   <dim:field mdschema="dc" element="contributor" qualifier="advisor" lang="en_US">Peter Szolovits.</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="author" lang="en_US">Zhang, Ying, 1976-</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="other" lang="en_US">Massachusetts Institute of Technology. Dept. of Electrical Engineering and Computer Science.</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="department">Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="accessioned">2005-06-02T19:40:53Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="available">2005-06-02T19:40:53Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="copyright" lang="en_US">2003</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="issued" lang="en_US">2003</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">http://hdl.handle.net/1721.1/18026</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="oclc" lang="en_US">57225475</dim:field>
   <dim:field mdschema="dc" element="description" lang="en_US">Thesis (M. Eng.)--Massachusetts Institute of Technology, Dept. of Electrical Engineering and Computer Science, 2003.</dim:field>
   <dim:field mdschema="dc" element="description" lang="en_US">Includes bibliographical references (p. 86-90).</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract" lang="en_US">The lack of effective data integration and knowledge representation in patient monitoring limits its utility to clinicians. Intelligent alarm algorithms that use artificial intelligence techniques have the potential to reduce false alarm rates and to improve data integration and knowledge representation. Crucial to the development of such algorithms is a well-annotated data set. In previous studies, clinical events were either unavailable or annotated without accurate time synchronization with physiological signals, generating uncertainties during both the development and evaluation of intelligent alarm algorithms. This research aims to help eliminate these uncertainties by designing a system that simultaneously collects physiological data and clinical annotations at the bedside, and to develop alarm algorithms in real time based on patient-specific data collected while using this system. In a standard pediatric intensive care unit, a working prototype of this system has helped collect a dataset of 196 hours of vital sign measurements at 1 Hz with 325 alarms generated by the bedside monitor and 2 instances of false negatives. About 89% of these alarms were clinically relevant true positives; 6% were true positives without clinical relevance; and 5% were false positives. Real-time machine learning showed improved performance over time and generated alarm algorithms that outperformed the previous generation of bedside monitors and came close in performance to the new generation. Results from this research suggest that the alarm algorithm(s) of the new patient monitoring systems have significantly improved sensitivity and specificity. They also demonstrated the feasibility of real-time learning at the bedside. Overall, they indicate</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract" lang="en_US">(cont.) that the methods developed in this research have the potential of helping provide patient-specific decision support for critical care.</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="statementofresponsibility" lang="en_US">b y Ying Zhang.</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degree" lang="en_US">M.Eng.</dim:field>
   <dim:field mdschema="dc" element="format" qualifier="extent" lang="en_US">94 p.</dim:field>
   <dim:field mdschema="dc" element="format" qualifier="extent">5118908 bytes</dim:field>
   <dim:field mdschema="dc" element="format" qualifier="extent">5129794 bytes</dim:field>
   <dim:field mdschema="dc" element="format" qualifier="mimetype">application/pdf</dim:field>
   <dim:field mdschema="dc" element="format" qualifier="mimetype">application/pdf</dim:field>
   <dim:field mdschema="dc" element="language" qualifier="iso" lang="en_US">eng</dim:field>
   <dim:field mdschema="dc" element="publisher" lang="en_US">Massachusetts Institute of Technology</dim:field>
   <dim:field mdschema="dc" element="rights" lang="en_US">M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.</dim:field>
   <dim:field mdschema="dc" element="rights" qualifier="uri">http://dspace.mit.edu/handle/1721.1/7582</dim:field>
   <dim:field mdschema="dc" element="subject" lang="en_US">Electrical Engineering and Computer Science.</dim:field>
   <dim:field mdschema="dc" element="title" lang="en_US">Real-time analysis of physiological data and development of alarm algorithms for patient monitoring in the Intensive Care Unit</dim:field>
   <dim:field mdschema="dc" element="type" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="dc" element="format" qualifier="mimetype">application/pdf</dim:field>
   <dim:field mdschema="dspace" element="authorsordered">false</dim:field>
   <dim:field mdschema="dspace" element="entity" qualifier="type">Publication</dim:field>
   <dim:field mdschema="others" element="access-status">unknown</dim:field>
   <dim:field mdschema="others" element="access-status">unknown</dim:field>
   <dim:field mdschema="cerif" element="openaire" authority="" confidence="-1">&lt;Publication xmlns="https://www.openaire.eu/cerif-profile/1.1/" id="703f1767-c778-40d3-96a4-48a80353309a">
	&lt;Type xmlns="https://www.openaire.eu/cerif-profile/vocab/COAR_Publication_Types">http://purl.org/coar/resource_type/c_1843&lt;/Type>
	&lt;Language>eng&lt;/Language>
   	&lt;Title>Real-time analysis of physiological data and development of alarm algorithms for patient monitoring in the Intensive Care Unit&lt;/Title>
   	&lt;PublishedIn>
    	&lt;Publication>
      	&lt;/Publication>
   	&lt;/PublishedIn>
   	&lt;PublicationDate>2003&lt;/PublicationDate>
   	&lt;Authors>
      	&lt;Author>
        	&lt;DisplayName>Zhang, Ying, 1976-&lt;/DisplayName>
         	&lt;Affiliation>
         		&lt;OrgUnit>
         		&lt;/OrgUnit>
         	&lt;/Affiliation>
      	&lt;/Author>
	&lt;/Authors>
   	&lt;Editors>
	&lt;/Editors>
    &lt;Publishers>
        &lt;Publisher>
            &lt;DisplayName>Massachusetts Institute of Technology&lt;/DisplayName>
            &lt;OrgUnit />
        &lt;/Publisher>
    &lt;/Publishers>
    &lt;License>http://dspace.mit.edu/handle/1721.1/7582&lt;/License>
    &lt;Keyword>Electrical Engineering and Computer Science.&lt;/Keyword>
   	&lt;Abstract>The lack of effective data integration and knowledge representation in patient monitoring limits its utility to clinicians. Intelligent alarm algorithms that use artificial intelligence techniques have the potential to reduce false alarm rates and to improve data integration and knowledge representation. Crucial to the development of such algorithms is a well-annotated data set. In previous studies, clinical events were either unavailable or annotated without accurate time synchronization with physiological signals, generating uncertainties during both the development and evaluation of intelligent alarm algorithms. This research aims to help eliminate these uncertainties by designing a system that simultaneously collects physiological data and clinical annotations at the bedside, and to develop alarm algorithms in real time based on patient-specific data collected while using this system. In a standard pediatric intensive care unit, a working prototype of this system has helped collect a dataset of 196 hours of vital sign measurements at 1 Hz with 325 alarms generated by the bedside monitor and 2 instances of false negatives. About 89% of these alarms were clinically relevant true positives; 6% were true positives without clinical relevance; and 5% were false positives. Real-time machine learning showed improved performance over time and generated alarm algorithms that outperformed the previous generation of bedside monitors and came close in performance to the new generation. Results from this research suggest that the alarm algorithm(s) of the new patient monitoring systems have significantly improved sensitivity and specificity. They also demonstrated the feasibility of real-time learning at the bedside. Overall, they indicate&lt;/Abstract>
   	&lt;Abstract>(cont.) that the methods developed in this research have the potential of helping provide patient-specific decision support for critical care.&lt;/Abstract>
	&lt;Access xmlns="http://purl.org/coar/access_right" 
    >
    &lt;/Access>
&lt;/Publication>
</dim:field>
</dim:dim>