dc.description.abstractPhase Contrast Imaging (PCI) was studied with the goal of understanding its relevance and its requirements. Current literature does not provide insight on the effect of a relaxation in coherence requirements on the PCI capabilities of an imaging system. This problem is all the more important since coherent X-ray and Neutron sources are mostly unavailable. We develop a model for PCI contribution to imaging for partially incoherent systems, and develop a methodology to identify a minimum and an optimum coherence length 4min and opt. We propose a figure-of-merit KPcI that quantifies the PCI capabilities of an imaging system. Our calculations show that X-ray PCI systems based on free space propagation using microfocus X-ray tubes have little PCI capabilities. We develop a model to explain the edge enhancement observed with those systems; our results suggest that scatter reduction is the process responsible for the observed edge enhancement. We performed experiments that show good agreement with the model. Coded Source Imaging (CSI) is proposed as a tool to produce highly coherent sources. The general theory of CSI is developed. We propose two possible systems: Fluorescent Coded Sources (FCS) and the AEB Encoded X-ray tube.en_US
dc.description.statementofresponsibilityby Antonio Leonardo Damato.en_US
dc.format.extent167 p.en_US
dc.language.isoengen_US
dc.publisherMassachusetts Institute of Technologyen_US
dc.rightsM.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.en_US
dc.rights.urihttp://dspace.mit.edu/handle/1721.1/7582en_US
dc.subjectNuclear Science and Engineering.en_US
dc.titleCapabilities and limitations of Phase Contrast Imaging techniques with X-rays and neutronsen_US
dc.title.alternativeCapabilities and limitations of PCI techniques with X-rays and neutronsen_US
dc.typeThesisen_US
dc.description.degreePh.D.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Nuclear Science and Engineering
dc.identifier.oclc540833901en_US