False arrhythmia alarm suppression using ECG, ABP, and photoplethysmogram
Author(s)
Deshmane, Anagha Vishwas
DownloadFull printable version (712.1Kb)
Other Contributors
Massachusetts Institute of Technology. Dept. of Electrical Engineering and Computer Science.
Advisor
Roger G. Mark and Lauren J. Kessler.
Terms of use
Metadata
Show full item recordAbstract
A signal quality assessment scheme for the photoplethysmogram waveform recorded by a pulse oximeter has been created. The signal quality algorithm uses statistical methods on time-series and spectral analysis to locate high-frequency segments of the photoplethysmogram waveform. A photoplethysmogram pulse onset detector has been implemented for heart rate estimation. Application of the signal quality metric and photoplethysmogram pulse onset detector are demonstrated in an algorithm which suppresses false electrocardiogram critical arrhythmia alarms issued by bedside monitors in hospital intensive care units.
Description
Thesis (M. Eng.)--Massachusetts Institute of Technology, Dept. of Electrical Engineering and Computer Science, 2009. This electronic version was submitted by the student author. The certified thesis is available in the Institute Archives and Special Collections. Cataloged from student-submitted PDF version of thesis. Includes bibliographical references (p. 91-93).
Date issued
2009Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer SciencePublisher
Massachusetts Institute of Technology
Keywords
Electrical Engineering and Computer Science.