dc.description.abstractPurpose We compare the performance of eight parallel transmit (pTx) body arrays with up to 32 channels and a standard birdcage design. Excitation uniformity, local specific absorption rate (SAR), global SAR, and power metrics are analyzed in the torso at 3 T for radiofrequency (RF)-shimming and 2-spoke excitations. Methods We used a fast cosimulation strategy for field calculation in the presence of coupling between transmit channels. We designed spoke pulses using magnitude least squares optimization with explicit constraint of SAR and power and compared the performance of the different pTx coils using the L-curve method. Results PTx arrays outperformed the conventional birdcage coil in all metrics except peak and average power efficiency. The presence of coupling exacerbated this power efficiency problem. At constant excitation fidelity, the pTx array with 24 channels arranged in three z-rows could decrease local SAR more than 4-fold (2-fold) for RF-shimming (2-spoke) compared to the birdcage coil for pulses of equal duration. Multi-row pTx coils had a marked performance advantage compared to single row designs, especially for coronal imaging. Conclusion PTx coils can simultaneously improve the excitation uniformity and reduce SAR compared to a birdcage coil when SAR metrics are explicitly constrained in the pulse design.en_US
dc.description.sponsorshipNational Institutes of Health (U.S.) (R01EB0068547)en_US
dc.description.sponsorshipNational Institutes of Health (U.S.) (R01EB007942)en_US
dc.description.sponsorshipNational Institutes of Health (U.S.) (P41EB015896)en_US
dc.language.isoen_US
dc.publisherWiley Blackwellen_US
dc.relation.isversionofhttp://dx.doi.org/10.1002/mrm.25243en_US
dc.rightsCreative Commons Attribution-Noncommercial-Share Alikeen_US
dc.rights.urihttp://creativecommons.org/licenses/by-nc-sa/4.0/en_US
dc.sourcePMCen_US
dc.titleComparison of simulated parallel transmit body arrays at 3 T using excitation uniformity, global SAR, local SAR, and power efficiency metricsen_US
dc.typeArticleen_US
dc.identifier.citationGuérin, Bastien; Gebhardt, Matthias; Serano, Peter et al. “Comparison of Simulated Parallel Transmit Body Arrays at 3 T Using Excitation Uniformity, Global SAR, Local SAR, and Power Efficiency Metrics.” Magnetic Resonance in Medicine 73, 3 (April 2014): 1137–1150 © 2014 Wiley Periodicals, Inc.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Electrical Engineering and Computer Scienceen_US
dc.contributor.mitauthorAdalsteinsson, Elfar
dc.relation.journalMagnetic Resonance in Medicineen_US
dc.eprint.versionAuthor's final manuscripten_US
dc.type.urihttp://purl.org/eprint/type/JournalArticleen_US
eprint.statushttp://purl.org/eprint/status/PeerRevieweden_US
dspace.orderedauthorsGuérin, Bastien; Gebhardt, Matthias; Serano, Peter; Adalsteinsson, Elfar; Hamm, Michael; Pfeuffer, Josef; Nistler, Juergen; Wald, Lawrence L.en_US
dspace.embargo.termsNen_US
dc.identifier.orcidhttps://orcid.org/0000-0002-7637-2914
mit.licenseOPEN_ACCESS_POLICYen_US
mit.metadata.statusComplete


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record