Testing and characterization of the TESS CCDs

Author(s)
Thayer, Carolyn T.; Villasenor, Jesus Noel Samonte; Morgan, Edward H; Prigozhin, Gregory; Prigozhin, Ilya; Ricker, George R; Sauerwein, Timothy A; Suntharalingam, Vyshnavi; Vanderspek, Roland K; Woods, Deborah F.; Kissel, Steve E.; La Marr, Beverly J.; ... Show more Show less
Thumbnail
Download99042X.pdf (2.359Mb)
PUBLISHER_POLICY

Publisher Policy

Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.

Terms of use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Metadata
Show full item record
Abstract
The Transiting Exoplanet Survey Satellite (TESS) is an Explorer-class mission dedicated to finding planets around bright, nearby stars so that more detailed follow-up studies can be done. TESS is due to launch in 2017 and careful characterization of the detectors will need to be completed on ground before then to ensure that the cameras will be within their photometric requirement of 60ppm/hr. TESS will fly MITLincoln Laboratories CCID-80s as the main scientific detector for its four cameras. They are 100μm deep depletion devices which have low dark current noise levels and can operate at low light levels at room temperature. They also each have a frame store region, which reduces smearing during readout and allows for near continuous integration. This paper describes the hardware and methodology that were developed for testing and characterizing individual CCID-80s. A dark system with no stimuli was used to measure the dark current. Fe 55 and Cd 109 X-ray sources were used to establish gain at low signal levels and its temperature dependence. An LED system that generates a programmable series of pulses was used in conjunction with an integrating sphere to measure pixel response non-uniformity (PRNU) and gain at higher signal levels. The same LED system was used with a pinhole system to evaluate the linearity and charge conservation capability of the CCID-80s.
Date issued
2016-07
URI
http://hdl.handle.net/1721.1/116358
Department
Lincoln Laboratory; MIT Kavli Institute for Astrophysics and Space Research
Journal
Proceedings of SPIE--the Society of Photo-Optical Instrumentation Engineers
Publisher
SPIE-Intl Soc Optical Eng
Citation
Thayer, C., J. Villasenor, S. Kissel, B. LaMarr, E. Morgan, G. Prigozhin, I. Prigozhin, et al. “Testing and Characterization of the TESS CCDs.” Edited by Howard A. MacEwen, Giovanni G. Fazio, Makenzie Lystrup, Natalie Batalha, Nicholas Siegler, and Edward C. Tong. Space Telescopes and Instrumentation 2016: Optical, Infrared, and Millimeter Wave (July 29, 2016).
Version: Final published version
ISSN
0277-786X
Keywords
CCD, dark current, gain, pixel response non-uniformity, linearity, TESS, exoplanets

Collections
  • MIT Open Access Articles

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.