Pseudo-Bayesian Model-Based Noninvasive Intracranial Pressure Estimation and Tracking

Author(s)
Imaduddin, Syed M; Fanelli, Andrea; Vonberg, Frederick W.; Tasker, Robert C.; Heldt, Thomas
Thumbnail
Download08836658.pdf (3.053Mb)
Publisher with Creative Commons License

Publisher with Creative Commons License

Creative Commons Attribution

Terms of use
Creative Commons Attribution 4.0 International license https://creativecommons.org/licenses/by/4.0/
Metadata
Show full item record
Abstract
Objective: A noninvasive intracranial pressure (ICP) estimation method is proposed that incorporates a model-based approach within a probabilistic framework to mitigate the effects of data and modeling uncertainties. Methods: A first-order model of the cerebral vasculature relates measured arterial blood pressure (ABP) and cerebral blood flow velocity (CBFV) to ICP. The model is driven by the ABP waveform and is solved for a range of mean ICP values to predict the CBFV waveform. The resulting errors between measured and predicted CBFV are transformed into likelihoods for each candidate ICP in two steps. First, a baseline ICP estimate is established over five data windows of 20 beats by combining the likelihoods with a prior distribution of the ICP to yield an a posteriori distribution whose median is taken as the baseline ICP estimate. A single-state model of cerebral autoregulatory dynamics is then employed in subsequent data windows to track changes in the baseline by combining ICP estimates obtained with a uniform prior belief and model-predicted ICP. For each data window, the estimated model parameters are also used to determine the ICP pulse pressure. Results: On a dataset of thirteen pediatric patients with a variety of pathological conditions requiring invasive ICP monitoring, the method yielded for mean ICP estimation a bias (mean error) of 0.6 mmHg and a root-mean-squared error of 3.7 mmHg. Conclusion: These performance characteristics are well within the acceptable range for clinical decision making. Significance: The method proposed here constitutes a significant step towards robust, continuous, patient-specific noninvasive ICP determination.
Date issued
2020-06
URI
https://hdl.handle.net/1721.1/130130
Department
Massachusetts Institute of Technology. Institute for Medical Engineering & Science; Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science; Massachusetts Institute of Technology. Research Laboratory of Electronics
Journal
IEEE Transactions on Biomedical Engineering
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Imaduddin, Syed M. et al. "Pseudo-Bayesian Model-Based Noninvasive Intracranial Pressure Estimation and Tracking." IEEE Transactions on Biomedical Engineering 67, 6 (June 2020): 1604 - 1615.
Version: Final published version
ISSN
0018-9294
1558-2531

Collections
  • MIT Open Access Articles

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.