In situ measurement of localization error in particle tracking microrheology

Author(s)
Josephson, Lilian L; Swan, James W; Furst, Eric M
Thumbnail
Download397_2018_1117_ReferencePDF.pdf (335.0Kb)
Open Access Policy

Open Access Policy

Creative Commons Attribution-Noncommercial-Share Alike

Terms of use
Creative Commons Attribution-Noncommercial-Share Alike http://creativecommons.org/licenses/by-nc-sa/4.0/
Metadata
Show full item record
Abstract
Abstract A method is presented to estimate the static error 𝜖 in a multiple particle tracking microrheology experiment. This in situ estimate of 𝜖 is measured under the same conditions of the material under test, and without any additional experiments. The correction of the mean-squared displacement by the in situ method is potentially more reliable than methods that rely on characterizing 𝜖 in a separate gel sample. With the true mean-squared displacements accessible at short lag times, experimental artifacts introduced by static error can be distinguished from true rheological properties, even in highly viscous (>10,000 mPa⋅s) samples.
Date issued
2018-10-27
URI
https://hdl.handle.net/1721.1/131424
Department
Massachusetts Institute of Technology. Department of Chemical Engineering
Publisher
Springer Berlin Heidelberg

Collections
  • MIT Open Access Articles

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.