dc.contributor.authorSalehi, Fatholah
dc.contributor.authorThome, Kurtis
dc.contributor.authorWenny, Brian N.
dc.contributor.authorLockwood, Ronald
dc.contributor.authorWang, Zhipeng
dc.date.accessioned2022-05-13T14:38:14Z
dc.date.available2022-05-13T14:38:14Z
dc.date.issued2022-05-10
dc.identifier.urihttps://hdl.handle.net/1721.1/142523
dc.description.abstractPrelaunch absolute, SI-traceable radiometric calibration of satellite-based sensors is key to ensuring the utility of imaging spectrometer-based data products. The development of detector-based calibration techniques leads to the feasibility of meeting the 0.3% uncertainty level needed to provide climate quality data sets. Detector-based calibration is a method in which a well-understood and stable transfer radiometer is calibrated in a standards laboratory to SI-traceable standards, and transported to a facility calibrating a sensor of interest. The transfer radiometer provides the calibration of the source used in the radiometric calibration. A detector-based calibration approach is part of the prelaunch calibration of the CLARREO (Climate Absolute Radiance and Refractivity Observatory) Pathfinder (CPF) sensor with the Goddard Laser for Absolute Measurement of Radiance (GLAMR) system. The SI-traceability of GLAMR is through the electric watt as part of the absolute radiometric calibration of the detectors at the National Institute of Standards and Technology using the Primary Optical Watt Radiometer. The current work uses GLAMR data collected with a visible and near-infrared imaging spectrometer calibration demonstration system to develop a source/sensor modeled calibration data set as part of a sensitivity study to evaluate uncertainties from the spectral sampling and processing methods that accompany the GLAMR calibration process. The spectral “supersets” include realistic instrumental features as well as effects from the GLAMR source. The methods needed to ensure that spurious sensor and GLAMR data are excluded are described. Results are given from the sensitivity study related to GLAMR spectral sampling and signal-to-noise ratio (SNR) effects, sensor integration time, and frame averaging of the imaging spectrometer data. The study shows that the 6 nm bandwidth sensor simulation requires a 1 nm spectral sampling of the GLAMR source with a radiance level that provides an in-band peak SNR > 200 to ensure that climate quality accuracies can be achieved. The results are also used to refine the test plan for the independent calibration for the CLARREO Pathfinder sensor calibration to optimize test time while meeting the required accuracy levels.en_US
dc.publisherMultidisciplinary Digital Publishing Instituteen_US
dc.relation.isversionofhttp://dx.doi.org/10.3390/rs14102302en_US
dc.rightsCreative Commons Attributionen_US
dc.rights.urihttps://creativecommons.org/licenses/by/4.0en_US
dc.sourceMultidisciplinary Digital Publishing Instituteen_US
dc.titleBand-Averaged Response Sensitivity Study of an Imaging Spectrometer for the CLARREO Pathfinder Missionen_US
dc.typeArticleen_US
dc.identifier.citationRemote Sensing 14 (10): 2302 (2022)en_US
dc.contributor.departmentLincoln Laboratory
dc.identifier.mitlicensePUBLISHER_CC
dc.eprint.versionFinal published versionen_US
dc.type.urihttp://purl.org/eprint/type/JournalArticleen_US
eprint.statushttp://purl.org/eprint/status/PeerRevieweden_US
dc.date.updated2022-05-12T19:36:25Z
dspace.date.submission2022-05-12T19:36:25Z
mit.licensePUBLISHER_CC
mit.metadata.statusAuthority Work and Publication Information Neededen_US