Increasing the Sensitivity of Enzyme-Linked Immunosorbent Assay Using Multiplexed Electrokinetic Concentrator
Author(s)
Cheow, Lih Feng; Ko, Sung Hee; Kim, Sung Jae; Kang, Kwan Hyoung; Han, Jongyoon
DownloadHan_Increasing the sensitivity.pdf (1.766Mb)
OPEN_ACCESS_POLICY
Open Access Policy
Creative Commons Attribution-Noncommercial-Share Alike
Alternative title
Increasing the Sensitivity of ELISA using Multiplexed
Electrokinetic Concentrator
Terms of use
Metadata
Show full item recordAbstract
We developed a novel method to increase the sensitivity of standard enzyme-linked immunosorbent assay (ELISA) using a multiplexed electrokinetic concentration chip. The poly(dimethylsiloxane) (PDMS) molecular concentrator(1) was used to trap and collect charged fluorescent product of target-bound enzyme turnover reaction of ELISA that occurred in a standard 96 well plate. Detection sensitivities of both prostate specific antigen (PSA) and CA 19-9 (a human pancreatic and gastrointestinal cancer marker) ELISAs in serum are enhanced ~100 fold with a low CV of <17%. We also integrated this method with an on-chip bead-based ELISA that lends itself toward a fully automated on-chip diagnostic device. Detection sensitivity of microfluidic bead-based CA 19-9 ELISA in serum is enhanced ~65 fold compared to the results without the electrokinetic accumulation step. This chip can be directly applied to enhance the readout sensitivity of a wide range of existing ELISA kits at concentrations below the current detection limit.
Date issued
2010-03Department
Massachusetts Institute of Technology. Department of Biological Engineering; Massachusetts Institute of Technology. Department of Electrical Engineering and Computer ScienceJournal
Analytical Chemistry
Publisher
American Chemical Society (ACS)
Citation
Cheow, Lih Feng et al. “Increasing the Sensitivity of Enzyme-Linked Immunosorbent Assay Using Multiplexed Electrokinetic Concentrator.” Analytical Chemistry 82.8 (2010): 3383–3388.
Version: Author's final manuscript
ISSN
0003-2700
1520-6882