Show simple item record

dc.contributor.authorChen, Yi
dc.contributor.authorTanaka, Shigehiko
dc.contributor.authorHunter, Ian
dc.date.accessioned2013-04-26T16:55:23Z
dc.date.available2013-04-26T16:55:23Z
dc.date.issued2010-08
dc.date.submitted2010-04
dc.identifier.isbn978-1-4244-4123-5
dc.identifier.issn1094-687X
dc.identifier.issn1558-4615
dc.identifier.otherINSPEC Accession Number: 11660502
dc.identifier.urihttp://hdl.handle.net/1721.1/78623
dc.description.abstractA disposable endoscopic platform with actuation motors inside the body of the endoscope is presented. This platform can enable new medical devices for diagnosis and for minimally invasive surgeries. This paper addresses mechanical and safety issues with existing endoscope technologies by incorporating disposability, safety modules, and lower cable forces. In order to produce path-independent cable forces, motors are incorporated in the body of the endoscope near the bending tip. Results for tip forces are shown accompanied by an analytical model describing the scaling laws for this type of robotic architecture. The system under development will provide a platform for research into haptic control and perceptual feedback.en_US
dc.description.sponsorshipNational Science Foundation (U.S.) (Fellowship)en_US
dc.language.isoen_US
dc.publisherInstitute of Electrical and Electronics Engineersen_US
dc.relation.isversionofhttp://dx.doi.org/10.1109/iembs.2010.5627677en_US
dc.rightsArticle is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.en_US
dc.sourceIEEEen_US
dc.titleDisposable endoscope tip actuation design and robotic platformen_US
dc.typeArticleen_US
dc.identifier.citationYi Chen, S Tanaka, and I W Hunter. "Disposable Endoscope Tip Actuation Design and Robotic Platform." 32nd Annual International Conference of the IEEE EMBS, Buenos Aires, Argentina, August 31-September 4, 2010, Pp. 2279–2282. IEEE. © Copyright 2010 IEEE.en_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Mechanical Engineeringen_US
dc.contributor.mitauthorHunter, Ian
dc.contributor.mitauthorTanaka, Shigehiko
dc.contributor.mitauthorChen, Yi
dc.relation.journalProceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2010en_US
dc.eprint.versionFinal published versionen_US
dc.identifier.pmid21097015
dc.type.urihttp://purl.org/eprint/type/ConferencePaperen_US
dspace.orderedauthorsYi Chen; Tanaka, S; Hunter, I Wen
dc.identifier.orcidhttps://orcid.org/0000-0003-2314-3031
dc.identifier.orcidhttps://orcid.org/0000-0002-8251-5432
mit.licensePUBLISHER_POLICYen_US
mit.metadata.statusComplete


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record