Lump detection with a gelsight sensor

Author(s)
Jia, Xiaodan; Li, Rui; Srinivasan, Mandayam A.; Adelson, Edward H.
Thumbnail
DownloadAdelson_Lump detection.pdf (856.6Kb)
OPEN_ACCESS_POLICY

Open Access Policy

Creative Commons Attribution-Noncommercial-Share Alike

Terms of use
Creative Commons Attribution-Noncommercial-Share Alike http://creativecommons.org/licenses/by-nc-sa/4.0/
Metadata
Show full item record
Abstract
A GelSight sensor is a tactile sensing device comprising a clear elastomeric pad covered with a reflective membrane, coupled with optics to measure the membrane's deformations. When the pad is pressed against an object's surface, the membrane changes shape in accord with mechanical and geometrical properties of the object. Since soft tissue is more compliant than hard tissue, one can detect an embedded lump by pressing the GelSight pad against the tissue surface and observing the hump that forms over the lump. We tested this system's sensitivity by constructing phantoms of soft rubber with hard embedded lumps. The system is quite sensitive; for example it could detect a 2mm lump at a depth of 5mm. The sensor was more sensitive than previous tactile lump detectors. It was also better than human observers using their fingertips. Such a capability could help in tumor screening, and could augment the sensory information available in telemedicine or minimally invasive surgery.
Date issued
2013-04
URI
http://hdl.handle.net/1721.1/86139
Department
Massachusetts Institute of Technology. Department of Brain and Cognitive Sciences; Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science; Massachusetts Institute of Technology. Department of Mechanical Engineering; Massachusetts Institute of Technology. Laboratory for Human and Machine Haptics
Journal
Proceedings of the 2013 World Haptics Conference (WHC)
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Xiaodan Jia, Rui Li, M. A. Srinivasan, and E. H. Adelson. “Lump Detection with a Gelsight Sensor.” 2013 World Haptics Conference (WHC) (n.d.).
Version: Author's final manuscript
ISBN
978-1-4799-0088-6
978-1-4799-0087-9

Collections
  • MIT Open Access Articles

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.