Show simple item record

dc.contributor.authorBrandt, Stephanie Akemi
dc.contributor.authorD'Alfonso, Mariarosaria
dc.contributor.authorHu, Miao
dc.contributor.authorKlute, Markus
dc.contributor.authorNiu, Xinmei
dc.contributor.authorCMS HCAL Collaboration
dc.date.accessioned2019-06-17T20:16:00Z
dc.date.available2019-06-17T20:16:00Z
dc.date.issued2017-12
dc.date.submitted2017-07
dc.identifier.issn1748-0221
dc.identifier.urihttps://hdl.handle.net/1721.1/121335
dc.description.abstractThe Phase I upgrade of the CMS Hadron Endcap Calorimeters consists of new photodetectors (Silicon Photomultipliers in place of Hybrid Photo-Diodes) and front-end electronics. The upgrade will eliminate the noise and the calibration drift of the Hybrid Photo-Diodes and enable the mitigation of the radiation damage of the scintillators and the wavelength shifting fibers with a larger spectral acceptance of the Silicon Photomultipliers. The upgrade also includes increased longitudinal segmentation of the calorimeter readout, which allows pile-up mitigation and recalibration due to depth-dependent radiation damage. As a realistic operational test, the responses of the Hadron Endcap Calorimeter wedges were calibrated with a 60 Co radioactive source with upgrade electronics. The test successfully established the procedure for future source calibrations of the Hadron Endcap Calorimeters. Here we describe the instrumentation details and the operational experiences related to the sourcing test.en_US
dc.language.isoen
dc.publisherIOP Publishingen_US
dc.relation.isversionofhttp://dx.doi.org/10.1088/1748-0221/12/12/P12034en_US
dc.rightsCreative Commons Attribution 3.0 unported licenseen_US
dc.rights.urihttps://creativecommons.org/licenses/by/3.0/en_US
dc.sourceIOP Publishingen_US
dc.titleRadioactive source calibration test of the CMS Hadron Endcap Calorimeter test wedge with Phase I upgrade electronicsen_US
dc.typeArticleen_US
dc.identifier.citationChatrchyan, S. et al. "Radioactive source calibration test of the CMS Hadron Endcap Calorimeter test wedge with Phase I upgrade electronics." Journal of Instrumentation 12 (December 2017): P12034 © 2017 CERN for the benefit of the CMS collaborationen_US
dc.contributor.departmentMassachusetts Institute of Technology. Department of Physicsen_US
dc.relation.journalJournal of Instrumentationen_US
dc.eprint.versionFinal published versionen_US
dc.type.urihttp://purl.org/eprint/type/JournalArticleen_US
eprint.statushttp://purl.org/eprint/status/PeerRevieweden_US
dc.date.updated2019-06-05T13:42:04Z
dspace.date.submission2019-06-05T13:42:05Z
mit.journal.volume12en_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record