A mechanical approach for smooth surface fitting to delineate vessel walls in optical coherence tomography images

Author(s)
Olender, Max; Athanasiou, Lambros S.; Hernández, José M. de la Torre; Ben-Assa, Eyal; Rikhtegar Nezami, Farhad; Edelman, Elazer R; ... Show more Show less
Thumbnail
DownloadAccepted version (4.050Mb)
Terms of use
Creative Commons Attribution-Noncommercial-Share Alike http://creativecommons.org/licenses/by-nc-sa/4.0/
Metadata
Show full item record
Abstract
Automated analysis of vascular imaging techniques is limited by the inability to precisely determine arterial borders. Intravascular optical coherence tomography (OCT) offers unprecedented detail of artery wall structure and composition, but does not provide consistent visibility of the outer border of the vessel due to the limited penetration depth. Existing interpolation and surface fitting methods prove insufficient to accurately fill the gaps between the irregularly spaced and sometimes unreliably identified visible segments of the vessel outer border. This paper describes an intuitive, efficient, and flexible new method of 3D surface fitting and smoothing suitable for this task. An anisotropic linear-elastic mesh is fit to irregularly spaced and uncertain data points corresponding to visible segments of vessel borders, enabling the fully automated delineation of the entire inner and outer borders of diseased vessels in OCT images for the first time. In a clinical dataset, the proposed smooth surface fitting approach had great agreement when compared with human annotations: areas differed by just 11 ± 11% (0.93 ± 0.84 mm2), with a coefficient of determination of 0.89. Overlapping and non-overlapping area ratios were 0.91 and 0.18, respectively, with a sensitivity of 90.8 and specificity of 99.0. This spring mesh method of contour fitting significantly outperformed all alternative surface fitting and interpolation approaches tested. The application of this promising proposed method is expected to enhance clinical intervention and translational research using OCT.
Date issued
2019-06
URI
https://hdl.handle.net/1721.1/126586
Department
Massachusetts Institute of Technology. Institute for Medical Engineering & Science; Massachusetts Institute of Technology. Department of Mechanical Engineering
Journal
IEEE Transactions on Medical Imaging
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Olender, Max L. et al. "A mechanical approach for smooth surface fitting to delineate vessel walls in optical coherence tomography images." IEEE Transactions on Medical Imaging 38, 6 (June 2019): 1384-1397 ©2019 Author(s)
Version: Author's final manuscript
ISSN
1558-254X
0278-0062

Collections
  • MIT Open Access Articles

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.