Laser calibration system for time of flight scintillator arrays
Author(s)
Denniston, AW; Segarra, Efrain Patrick; Schmidt, Axel; Beck, Arie; Beck, S. May-Tal; Cruz Torres, Reynier; Hauenstein, F.; Hrnjic, A.; Kutz, Tyler; Nambrath, Anjali; Pybus, J. R.; Toldeo, P.; Hen, Or; Weinstein, L.B.; Olivenboim, M.; Piazetzky, E.; Korover, I.; ... Show more Show less
DownloadAccepted version (5.361Mb)
Publisher with Creative Commons License
Publisher with Creative Commons License
Creative Commons Attribution
Terms of use
Metadata
Show full item recordAbstract
© 2020 Elsevier B.V. A laser calibration system was developed for monitoring and calibrating time of flight (TOF) scintillating detector arrays. The system includes setups for both small- and large-scale scintillator arrays. Following test-bench characterization, the laser system was recently commissioned in experimental Hall B at the Thomas Jefferson National Accelerator Facility for use on the new Backward Angle Neutron Detector (BAND) scintillator array. The system successfully provided time walk corrections, absolute time calibration, and TOF drift correction for the scintillators in BAND. This showcases the general applicability of the system for use on high-precision TOF detectors.
Date issued
2020Department
Massachusetts Institute of Technology. Department of Physics; Massachusetts Institute of Technology. Laboratory for Nuclear ScienceJournal
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Publisher
Elsevier BV